期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 877, 期 27, 页码 3215-3220出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2009.08.037
关键词
para-Aminohippuric acid; High-performance liquid chromatography; Mass spectrometry; Human plasma; Human urine
资金
- Pfizer Global RD
In this manuscript, we present a simple and reliable method for the quantitation of para-aminohippuric acid (PAH; 2-(4-aminobenzamido)acetic acid) in human plasma and urine using liquid chromatography coupled to electrospray ionisation low-energy collision-induced dissociation tandem mass spectrometry (HPLC-ESI-CID-MS/MS) analysis (negative ion mode) via multiple reaction monitoring (MRM). Sample preparation involved protein precipitation of plasma samples with acetonitrile and subsequent dilution of urine samples with the mobile phase. The internal standard (IS) adopted was para-aminosalicylic acid (PAS; 4-amino-2-hydroxybenzoic acid). Chromatographic separation was achieved on a Cosmosil HILIC column using an isocratic mobile phase consisting of ammonium acetate buffer (20 mM) and acetonitrile (45:55, v/v) at a flow rate of 200 mu l/min. The transactions monitored were m/z 192.9 -> 149.1 for PAH and m/z 152.1 -> 108.1 for IS. Linear calibration curves were generated over the PAH concentration range of 0.2-100 mg/L in human plasma and urine. The method was validated for selectivity, accuracy, precision, recovery and stability according to USFDA criteria, and has been successfully applied to a pharmacokinetic study in healthy volunteers administered an intravenous dose of 440 mg PAH. (C) 2009 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据