期刊
JOURNAL OF CHROMATOGRAPHY A
卷 1208, 期 1-2, 页码 71-75出版社
ELSEVIER
DOI: 10.1016/j.chroma.2008.08.091
关键词
Sample preparation; Monolith silica; Amphetamines; Methylenedioxyamphetamines
资金
- Japan Society for the Promotion of Science [15922167]
- Grants-in-Aid for Scientific Research [15922167] Funding Source: KAKEN
To overcome the limitations of solid-phase extraction, we developed a device comprising a spin column packed with octadecyl silane-bonded monolithic silica for extracting amphetamines and methylenedioxyamphetamines from urine. Urine (0.5 mL), buffer (0.4 mL), and methoxyphenamine (internal standard) were directly put into the preactivated column. The column was centrifuged (3000 rpm, 5 min) for sample loading and washed. The adsorbed analytes were eluted and analyzed by high-performance liquid chromatography, without evaporation. The results were as follows: linear curves (drug concentrations of 0.2-20 mu g/mL); correlation coefficients >0.99; detection limit, 0.1 mu g/mL The proposed method is not only useful for drugs from biological materials but also highly reproducible for the analysis of these drugs in urine. (C) 2008 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据