3.9 Article

Cementless total hip replacement using second-generation components A 12- TO 16-YEAR FOLLOW-UP

期刊

JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME
卷 92B, 期 12, 页码 1636-1641

出版社

BRITISH EDITORIAL SOC BONE JOINT SURGERY
DOI: 10.1302/0301-620X.92B12.24582

关键词

-

向作者/读者索取更多资源

We reviewed 123 second-generation uncemented total hip replacements performed on 115 patients by a single surgeon between 1993 and 1994. The acetabular component used in all cases was a fully porous-coated threaded hemispheric titanium shell (T-Tap ST) with a calcium ion stearate-free, isostatically compression-moulded polyethylene liner. The titanium femoral component used was a Taperloc with a reduced distal stem. No patient was lost to follow-up. Complete clinical and radiological follow-up was obtained for all 123 hips at a mean of 14 years (12 to 16). One femoral component was revised after a fracture, and three acetabular components for aseptic loosening. No additional femoral or acetabular components were judged loose by radiological criteria. Mild proximal femoral osteolysis was identified in two hips and minor acetabular osteolysis was present in four. The mean rate of penetration of the femoral head was 0.036 mm/year (0.000 to 0.227). These findings suggest that refinements in component design may be associated with excellent long-term fixation in cementless primary total hip replacement.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.9
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据