期刊
JOURNAL OF BIOPHOTONICS
卷 6, 期 6-7, 页码 549-559出版社
WILEY-V C H VERLAG GMBH
DOI: 10.1002/jbio.201200223
关键词
photoacoustic imaging; optoacoustic imaging; ultrasound imaging; section imaging; acoustic mirror; optical detection
A setup is proposed that provides perfectly co-registered photoacoustic (PA) and ultrasound (US) section images. Photoacoustic and ultrasound backscatter signals are generated by laser pulses coming from the same laser system, the latter by absorption of some of the laser energy on an optically absorbing target near the imaged object. By measuring both signals with the same optical detector, which is focused into the selected section by use of a cylindrical acoustic mirror, the information for both images is acquired simultaneously. Co-registered PA and US images are obtained after applying the inverse Radon transform to the data, which are gathered while rotating the object relative to the detector. Phantom experiments demonstrate a resolution of 1.1 mm between the sections of both imaging modalities and a in-plane resolution of about 60 mu m and 120 mu m for the US and PA modes, respectively. The complementary contrast mechanisms of the two modalities are shown by images of a zebrafish. ((c) 2013 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim)
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐