4.5 Article

Real-time monitoring of cardiac radio-frequency ablation lesion formation using an optical coherence tomography forward-imaging catheter

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 15, 期 3, 页码 -

出版社

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.3459134

关键词

radio-frequency ablation; optical coherence tomography; cardiac imaging; catheter

资金

  1. Wallace H. Coulter Foundation
  2. NIH [F31HL085939, R01HL083048, RR1246]

向作者/读者索取更多资源

Radio-frequency ablation (rfa) is the standard of care for the treatment of cardiac arrhythmias; however, there are no direct measures of the successful delivery of ablation lesions. Optical coherence tomography (OCT) imaging has the potential to provide real-time monitoring of cardiac rfa therapy, visualizing lesion formation and assessing tissue contact in the presence of blood. A rfa-compatible forward-imaging conical scanning probe is prototyped to meet this need. The forward-imaging probe provides circular scanning, with a 2-mm scan diameter and 30-mu m spot size. During the application of rf energy, dynamics are recorded at 20 frames per second with a 40-kHz A-line rate. Real-time monitoring of cardiac rfa lesion formation and imaging in the presence of blood is demonstrated ex vivo in a swine left ventricle with a forward, flexible, circular scanning OCT catheter. (C) 2010 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.3459134]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据