Ex vivo measurement of lumbar intervertebral disc pressure using fibre-Bragg gratings

期刊

JOURNAL OF BIOMECHANICS
卷 41, 期 1, 页码 221-225

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.jbiomech.2007.07.015

关键词

spine; intervertebral disc; low back pain; pressure; sensor; nucleus pulposus; disc degeneration

向作者/读者索取更多资源

Methods were developed to measure intervertebral disc pressure using optical fibre-Bragg gratings (FBGs). The FBG sensor was calibrated for hydrostatic pressure in a purpose-built apparatus and the average sensitivity was determined to be -5.7 +/- 0.085pm/MPa (mean +/- SD). The average coefficient of determination (r(2)) for the calibration data was 0.99, and the average hysteresis of the sensor was 2.13% of full scale. The FBG was used to measure intradiscal pressure response to compressive load in five lumbar functional spine units. The pressure measured by the FBG sensor varied linearly with applied compressive load with coefficients of determination ranging from 0.84 to 0.97. The FBG sensor's sensitivity to compressive load ranged from 0.702 +/- 0.043kPa/N (mean +/- SD) in a L1-L2 specimen, to 1.07 +/- 0.069 kPa/N in a L4-L5 specimen. These measurements agree with those of previous studies in lumbar spines. Two strain gauge pressure sensors were also used to measure intradiscal pressure response to compressive load. The measured pressure sensitivity to load ranged from 0.251 kPa/N (L4-L5) to 0.850kPa/N (L2-L3). The average difference in pressure sensitivity to load between Sensors 1 and 2 was 12.9% of the value for Sensor 1, with a range from 1.1% to 20.4%, which suggests that disc pressure was not purely hydrostatic. This may have contributed to the difference between the responses of the FBG and strain gauge sensors. (C) 2007 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据