4.7 Article

Development of microstretching device to evaluate cell membrane strain field around sensing point of mechanical stimuli

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.ijmecsci.2009.09.021

关键词

Biomechanics; Mechanotransduction; MEMS; Cellular response; Tensile test

资金

  1. MEXT (The Ministry of Education, Culture, Sports, Science and Technology) [21760081]
  2. Grants-in-Aid for Scientific Research [21760081] Funding Source: KAKEN

向作者/读者索取更多资源

In this study, we developed a novel cell-stretching device to evaluate the micro-strain field on a cell membrane in which a cellular response to a mechanical stimulus occurred. In a conventional cell-stretching device, rigid displacement of the cell body during the stretching operation is not negligible. This rigid displacement makes it difficult to observe the initial cellular response to stretching with high spatial and temporal resolution. A novel device is fabricated by applying the MEMS (micro-electromechanical systems) technique, and consists of a transparent elastic microchamber and a microlinkage mechanism by which uniaxial stretch is applied to the microchamber. The microchamber and linkage mechanism are fabricated by micromolding and multiple exposure of photoresist, respectively. We demonstrated the applicability of the fabrication process that we proposed, and evaluated the performance of the microdevice prototype. As a result of operation tests, rigid displacement of the cell body during stretching was found to be well suppressed. This result indicated the potential of this device to be used to observe and evaluate the initial cellular response and micro-strain field on a cell membrane during uniaxial stretching. (C) 2009 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据