Differential mobility spectrometry/mass spectrometry: The design of a new mass spectrometer for real-time chemical analysis in the field

期刊

INTERNATIONAL JOURNAL OF MASS SPECTROMETRY
卷 295, 期 3, 页码 138-144

出版社

ELSEVIER
DOI: 10.1016/j.ijms.2010.03.011

关键词

Portable mass spectrometer; Differential mobility spectrometry; Miniature ion funnel

资金

  1. U.S. Department of Energy [DE-AC52-06NA25946]
  2. Columbia University Center for Medical Countermeasures against Radiation
  3. NIH/NIAID [U19 AI067773-04]

向作者/读者索取更多资源

The design of a prototype, field-portable mass spectrometer (MS) is described. The MS has been designed with an atmospheric interface in order to couple the system to a commercially available differential mobility spectrometer. The differential mobility spectrometer provides selective injection of trace-level analytes of interest into the inlet of the MS for real-time chemical detection. To accomplish this task, the MS design incorporates the use of an electrodynamic ion funnel to transport the ion beam generated at atmospheric pressure, to the high-vacuum chamber that houses the mass analyzer. This leads to a design that utilizes two stages of differential pumping to achieve an overall pressure drop from atmosphere (760Torr) to approximately 1 x 10(-5)Torr. A mass resolution of 140 (m/z =84) and a limit of detection of approximately 1 ppb(v) have been measured for the system. (c) 2010 Elsevier B.V. All rights reserved.