4.4 Article

Evaluating the performance of the NPL femtosecond frequency combs: agreement at the 10-21 level

期刊

METROLOGIA
卷 52, 期 1, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0026-1394/52/1/62

关键词

frequency comb; optical frequency metrology; absolute frequency; optical frequency ratio

资金

  1. UK National Measurement System as part of the Electromagnetics and Time Programme

向作者/读者索取更多资源

Results are presented from a series of comparisons between two independent femtosecond frequency comb systems at NPL, which were carried out in order to assess their systematic uncertainty. Simultaneous measurements with the two systems demonstrate agreement at the level of 5 x 10(-18) when measuring an optical frequency against a common microwave reference. When simultaneously measuring the ratio of two optical frequencies, agreement at the 3 x 10(-21) level is observed. The results represent the highest reported level of agreement to date between Ti: Sapphire and Er-doped femtosecond combs. The limitations of the combs when operating in these two different manners are discussed, including traceability to the SI second, which can be achieved with an uncertainty below 1 x 10(-16). The technical details presented underpin recent absolute frequency measurements of the Sr-88(+) and Yb-171(+) optical clock transitions at NPL, as well as a frequency ratio measurement between the two optical clock transitions in Yb-171(+).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据