Development of a Wireless Sensor for the Measurement of Chicken Blood Flow Using the Laser Doppler Blood Flow Meter Technique

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 60, 期 6, 页码 1645-1653

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2013.2241062

关键词

Avian influenza; blood flow; laser Doppler; low power consumption; microsensor; wireless

资金

  1. Japan Science and Technology Agency

向作者/读者索取更多资源

Here, we report the development of an integrated laser Doppler blood flow micrometer for chickens. This sensor weighs only 18 g and is one of the smallest-sized blood flow meters, with no wired line, these are features necessary for attaching the sensor to the chicken. The structure of the sensor chip consists of two silicon cavities with a photo diode and a laser diode, which was achieved using the microelectromechanical systems technique, resulting in its small size and significantly low power consumption. In addition, we introduced an intermittent measuring arrangement in the measuring system to reduce power consumption and to enable the sensor to work longer. We were successfully able to measure chicken blood flow for five consecutive days, and discovered that chicken blood flow shows daily fluctuations.