期刊
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY
卷 20, 期 3, 页码 818-821出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TASC.2009.2039014
关键词
Finite element method; HTS RF coil; MRI; signal-to-noise ratio
A Bi-2223 surface RF coil was developed for a 0.3T permanent MRI system to achieve better signal-to-noise ratio (SNR) of images. Based on the finite element method (FEM), the SNR of a copper-made surface coil is analyzed, and the emphasis is put on the coil self-resistance that accounts for much higher percentage of the equivalent noise resistance in low field MRI than that in high field MRI. In order to improve SNR, a surface coil that is made of a Bi-2223 tape is fabricated for a 0.3T permanent MRI system. Phantom images are acquired, and the image SNR by the Bi-2223 coil in liquid nitrogen is improved by up to 35% compared with that by a copper coil at room temperature.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐