4.4 Article Proceedings Paper

Signal-to-Noise Ratio Improvement by Bi-2223 Surface RF Coil in 0.3T MRI System

期刊

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TASC.2009.2039014

关键词

Finite element method; HTS RF coil; MRI; signal-to-noise ratio

向作者/读者索取更多资源

A Bi-2223 surface RF coil was developed for a 0.3T permanent MRI system to achieve better signal-to-noise ratio (SNR) of images. Based on the finite element method (FEM), the SNR of a copper-made surface coil is analyzed, and the emphasis is put on the coil self-resistance that accounts for much higher percentage of the equivalent noise resistance in low field MRI than that in high field MRI. In order to improve SNR, a surface coil that is made of a Bi-2223 tape is fabricated for a 0.3T permanent MRI system. Phantom images are acquired, and the image SNR by the Bi-2223 coil in liquid nitrogen is improved by up to 35% compared with that by a copper coil at room temperature.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据