4.7 Article

Imaging of Breast Cancer With Optical Coherence Tomography Needle Probes: Feasibility and Initial Results

期刊

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSTQE.2011.2166757

关键词

Cancer; interstitial imaging; needle probe; OCT needle; optical coherence tomography (OCT)

资金

  1. National Breast Cancer Foundation, Sydney, Australia
  2. Cancer Council Western Australia, WA, Australia
  3. National Breast Cancer Foundation [NC-12-19] Funding Source: researchfish

向作者/读者索取更多资源

Optical coherence tomography (OCT) is a high-resolution imaging modality with the potential to provide in situ assessment to distinguish normal from cancerous tissue. However, limited image penetration depth has restricted its utility. This paper demonstrates the feasibility of an OCT needle probe to perform interstitial imaging deep below the tissue surface. The side-facing needle probe comprises miniaturized focusing optics consisting of no-core and GRIN fiber encased within either a 22- or 23-gauge needle. 3-D OCT volumetric data sets were acquired by rotating and retracting the probe during imaging. We present the first published image of a human breast cancer tumor margin, and of human axillary lymph nodes acquired with an OCT needle probe. Through accurate correlation with the histological gold standard, OCT is shown to enable a clear delineation of tumor boundary from surrounding adipose tissue, and identification of microarchitectural features.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据